ERIC™ Electronic Compendium (version 1.0.771) Patch 0613 for Thermo Scientific™ RapID™ Systems

Issue Date: August 14, 2013

Versions: ERIC™ 1.0.771


Condition: An ERIC software patch update ERICPatch0613 has been issued to modify and update ERIC 1.0.771 to accommodate database changes since the release of ERIC 1.0.771. The patch should not be applied to ERIC versions earlier than 1.0.771.


Burkholderia pseudomallei may be misidentified (most typically as Burkholderia cepacia) by the RapID NF Plus system. This patch will modify the database to improve identification of Burkholderia pseudomallei.


Recent taxonomic changes are not reflected in the ERIC software databases. This patch will modify the database with key taxonomic changes.


Solution: A software patch ERICPatch0613 has been issued to:

  • Update the RapID NF Plus database with revised probabilities for Burkholderia pseudomallei
  • Update organism names and associated notes in the following databases, where taxonomy has been revised:
    • RapID ANA II
    • RapID NH
    • RapID ONE
    • RapID SS/u
    • RapID STR
  • Address any typographical errors found in current notes.


The patch is a self-extracting application that will copy an executable ERICPatch0613.exe to the folder on the target computer that contains the ERIC database.


Instructions

Download ERIC Patch 0613 to obtain the file ERICPatch0613.zip from Remel. Click here to download the white paper (PDF document).

(Click here to go to 7-Zip web site to download the tool to open ZIP file, if required)

Or contact:

Technical Service

Phone: (800) 255-6730

Fax: (877) 428-1924

E-Mail: Technical Support


Step 1. Copy the file, ERICPatch0613.zip, to your computer. The file may be copied to any location, including your desktop. The file will contain both the self-extracting patch file and a copy of this set of instructions.


Step 2. Once copied to a location, open ERICPatch0613.zip and extract all files, including ERICPatch0613.exe. Double-click on the ERICPatch0613.exe file to launch the patch. The patch should be titled ERIC Patch 0613. Read the instructions on the screen and click "Next".


Step 3. You may be asked for the folder to which the file should be extracted. You must extract the files to the location on your computer containing ERIC.exe. Typically, ERIC is installed to the default location - C:\Program Files\Remel\ERIC (or C:\Program Files (x86)\Remel\ERIC for 64-bit Windows operating systems). If you selected the ERIC default at install, click "Next". If this is NOT the folder you selected when installing ERIC, please supply the location of the ERIC.EXE file folder and then click "Next".


Step 4. The patch will connect to ERIC and display a dialog box. Click "Apply" to apply the patch. The patch will display a message "Patch successfully applied" when complete. Exit the patch.


Step 5. Launch ERIC. Go to the Icon Bar, select "Control Panel" and then select "Updates". The patch applied should be listed in the Upgrade/Patch list.


Further assistance: For additional assistance, contact Thermo Scientific Technical Service.



Click here for the summary of taxonomic changes.